1. Airflow resistance test
2. Adsorption efficiency/penetration rate
3. Determine desorption
The main purpose
The mask filtration efficiency test device is used to test the protection effect and filtration efficiency of daily protective masks and respirators against particulates;
According to the standard
GB 32610-2016 "Technical Specification for Daily Protective Masks"
GB 2626-2006 "Respiratory protective equipment self-priming filter anti-particulate respirator"
GB 19081-2009 "Technical Requirements for Medical Disposable Protective Clothing"
GB 19083-2010 "Technical Requirements for Medical Protective Masks"
YY 0469 medical surgical mask
Technical Parameters
product name |
F9320I |
F9320 Ⅱ |
|
medium |
DOP,PAO-4 |
NaCl |
|
shape |
Size and distribution |
0.18 count median particle size; concentrated particle size 0.3μm |
0.075 count median particle size; concentrated particle size 0.26μm |
weight |
80kg |
||
aerosol |
The scope of work |
0%RH~100%RH (no condensation) |
|
standard deviation |
<1.6 |
<1.86 |
|
concentration |
80-120mg/m3 |
20-50mg/m3 |
|
Measurement/operating conditions |
effectiveness |
Up to 99.9995% |
|
pressure |
0~2500Pa, accuracy: ±2.5% of full scale |
||
flow |
10~110L/min, accuracy: ±(0.4% reading + 0.2% full scale) |
||
power supply |
220VAC/50Hz |
||
Compressed air |
283lpm@5.6kg/cm2 |
Fourth, the scope of application
Applicable to medical device inspection centers, safety protection inspection centers, labor protection inspection centers, drug inspection centers, disease prevention and control centers, hospitals, masks and respirator manufacturers, etc.;