Dedicated to innovation and application of testing equipment in environmental measurement, fluid measurement, and microparticle measurement.
+
  • 44de0cb2-aa59-46fd-bf83-28bbbeb1afdb.jpg

F9100 Series Air Filter Performance Test Equipment

1. Uses internationally advanced testing technology - sonic nozzle technology, which is internationally leading.
● High accuracy, good repeatability
● Rapid response, high sensitivity
● Stable performance, long verification cycle (5 years)
2. Adopts an automatic quantitative ash-adding system, eliminating human interference factors and ensuring the correctness of the test results.
● Uniform ash addition rate
● Accurate total ash addition - electronic balance weighing
3. Uses an industrial control computer and advanced control methods to achieve automation of the test process. Easy to operate.

Service Support:

  • Product Description
  • Meet the following standards

    (1) QC/T 32-2006

    (2) QC/T 970-2014

    (3) QC/T 793-2007

    (4) SAE J726-2002

    (5)ISO 5011:2000(E)

    (6)JB/T 9747-2005

    (7) JIS D1612

     

    Technical indicators

    Airflow Airflow range: (30~3000)m³/h (under standard conditions)
    Airflow accuracy: ±0.5%
    Upstream nozzle stagnation pressure Measurement range: (0~110)kPa
    Accuracy: ±0.1%
    Upstream nozzle stagnation temperature Measurement range: (0~50)℃

    Accuracy: Class A
    Ash feeding device Ash feeding range: (0.5~100)g/min
    Uniformity: The change in ash feeding amount every 5 minutes does not exceed 5% of the average ash feeding amount.
    Static pressure and pressure drop Measurement range: (0.1~20)kPa
    Accuracy: ±0.1% FS
    Measurement range: (0~6)kPa
    Accuracy: ±0.1% FS
    Static pressure measurement accuracy: ±0.1% FS
    Environmental condition measurement Temperature measurement
    Measurement range: (-20~80)℃
    Accuracy: ±0.5℃
    Humidity measurement
    Measurement range: (0~100)% RH
    Accuracy: ±2%RH
    Atmospheric pressure measurement accuracy ≤ ±0.3kpa
    Absolute filter mass measurement Measurement range: Electronic balance range is 2Kg
    Accuracy: 0.01g
    Dust injector Meets QC/T32 standard requirements

     